| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.079 | 1.995 |
| 1000.00 | 1.631 | 1.547 |
| 500.00 | 1.140 | 1.056 |
| 250.00 | 0.902 | 0.818 |
| 125.00 | 0.447 | 0.363 |
| 62.50 | 0.288 | 0.204 |
| 31.25 | 0.217 | 0.133 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-115% | 103% |
| EDTA plasma(n=5) | 84-97% | 91% |
| Heparin plasma(n=5) | 87-117% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-101% | 81-100% | 84-114% | 92-104% |
| EDTA plasma(n=5) | 86-92% | 91-103% | 89-113% | 91-100% |
| Heparin plasma(n=5) | 95-105% | 89-110% | 81-104% | 86-112% |