| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.082 | 1.998 |
| 5.00 | 1.656 | 1.572 |
| 2.50 | 1.162 | 1.078 |
| 1.25 | 0.811 | 0.727 |
| 0.63 | 0.530 | 0.446 |
| 0.31 | 0.349 | 0.265 |
| 0.16 | 0.242 | 0.158 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-107% | 98% |
| EDTA plasma(n=5) | 84-115% | 100% |
| Heparin plasma(n=5) | 82-107% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 87-119% | 86-96% | 81-97% |
| EDTA plasma(n=5) | 81-112% | 94-108% | 87-92% | 87-119% |
| Heparin plasma(n=5) | 94-118% | 90-104% | 93-100% | 85-118% |