| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.982 | 1.897 |
| 2500.00 | 1.589 | 1.504 |
| 1250.00 | 1.128 | 1.043 |
| 625.00 | 0.832 | 0.747 |
| 312.50 | 0.541 | 0.456 |
| 156.25 | 0.321 | 0.236 |
| 78.13 | 0.155 | 0.070 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-117% | 103% |
| EDTA plasma(n=5) | 90-96% | 93% |
| Heparin plasma(n=5) | 86-107% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-89% | 92-105% | 91-110% | 88-109% |
| EDTA plasma(n=5) | 90-100% | 89-107% | 88-100% | 82-115% |
| Heparin plasma(n=5) | 91-103% | 90-102% | 85-114% | 88-102% |
