| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.126 | 2.040 | 
| 500.00 | 1.593 | 1.507 | 
| 250.00 | 1.147 | 1.061 | 
| 125.00 | 0.871 | 0.785 | 
| 62.50 | 0.595 | 0.509 | 
| 31.25 | 0.318 | 0.232 | 
| 15.63 | 0.235 | 0.149 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 94-107% | 100% | 
| EDTA plasma(n=5) | 78-92% | 85% | 
| Heparin plasma(n=5) | 82-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 92-105% | 82-92% | 86-94% | 81-95% | 
| EDTA plasma(n=5) | 92-105% | 92-101% | 80-94% | 86-101% | 
| Heparin plasma(n=5) | 85-95% | 79-96% | 89-101% | 85-97% |