| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.227 | 2.143 |
| 5.00 | 1.690 | 1.606 |
| 2.50 | 1.126 | 1.042 |
| 1.25 | 0.850 | 0.766 |
| 0.63 | 0.501 | 0.417 |
| 0.31 | 0.323 | 0.239 |
| 0.16 | 0.190 | 0.106 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-94% | 88% |
| EDTA plasma(n=5) | 91-99% | 95% |
| Heparin plasma(n=5) | 94-99% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-108% | 86-115% | 85-107% | 90-108% |
| EDTA plasma(n=5) | 84-90% | 84-116% | 86-116% | 92-103% |
| Heparin plasma(n=5) | 91-116% | 94-119% | 86-117% | 89-99% |