| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.842 | 1.745 |
| 5.00 | 1.593 | 1.496 |
| 2.50 | 1.183 | 1.086 |
| 1.25 | 0.752 | 0.655 |
| 0.63 | 0.457 | 0.360 |
| 0.32 | 0.319 | 0.222 |
| 0.16 | 0.213 | 0.116 |
| 0.00 | 0.097 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-97% | 88% |
| EDTA plasma(n=5) | 83-95% | 89% |
| Heparin plasma(n=5) | 88-102% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-97% | 92-103% | 86-97% | 93-101% |
| EDTA plasma(n=5) | 87-96% | 86-95% | 91-103% | 87-93% |
| Heparin plasma(n=5) | 82-94% | 85-97% | 91-102% | 81-90% |