| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 4000.00 | 2.268 | 2.180 | 
| 2000.00 | 1.790 | 1.702 | 
| 1000.00 | 1.240 | 1.152 | 
| 500.00 | 0.882 | 0.794 | 
| 250.00 | 0.492 | 0.404 | 
| 125.00 | 0.262 | 0.174 | 
| 62.50 | 0.186 | 0.098 | 
| 0.00 | 0.088 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 85-97% | 91% | 
| EDTA plasma(n=5) | 83-99% | 95% | 
| Heparin plasma(n=5) | 85-97% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-92% | 89-96% | 90-98% | 87-98% | 
| EDTA plasma(n=5) | 88-99% | 86-92% | 89-103% | 97-105% | 
| Heparin plasma(n=5) | 95-104% | 87-96% | 82-96% | 89-103% |