| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 1.994 | 1.909 | 
| 5.00 | 1.565 | 1.480 | 
| 2.50 | 1.142 | 1.057 | 
| 1.25 | 0.897 | 0.812 | 
| 0.63 | 0.538 | 0.453 | 
| 0.32 | 0.389 | 0.304 | 
| 0.16 | 0.205 | 0.120 | 
| 0.00 | 0.085 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 88-107% | 94% | 
| EDTA plasma(n=5) | 97-105% | 101% | 
| Heparin plasma(n=5) | 87-99% | 92% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 91-98% | 83-101% | 87-98% | 95-102% | 
| EDTA plasma(n=5) | 88-97% | 78-96% | 80-99% | 89-97% | 
| Heparin plasma(n=5) | 85-101% | 87-103% | 84-92% | 78-90% |