| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.154 | 2.059 | 
| 500.00 | 1.717 | 1.622 | 
| 250.00 | 1.035 | 0.940 | 
| 125.00 | 0.828 | 0.733 | 
| 62.50 | 0.496 | 0.401 | 
| 31.25 | 0.315 | 0.220 | 
| 15.63 | 0.256 | 0.161 | 
| 0.00 | 0.095 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 95-107% | 101% | 
| EDTA plasma(n=5) | 80-93% | 86% | 
| Heparin plasma(n=5) | 94-105% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-102% | 79-94% | 83-96% | 85-106% | 
| EDTA plasma(n=5) | 88-97% | 90-99% | 85-97% | 79-101% | 
| Heparin plasma(n=5) | 78-94% | 97-105% | 88-101% | 82-96% |