| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (µg/mL) | OD | Corrected OD | 
|---|---|---|
| 60.00 | 2.207 | 2.122 | 
| 30.00 | 1.657 | 1.572 | 
| 15.00 | 1.196 | 1.111 | 
| 7.50 | 0.882 | 0.797 | 
| 3.75 | 0.576 | 0.491 | 
| 1.88 | 0.366 | 0.281 | 
| 0.94 | 0.213 | 0.128 | 
| 0.00 | 0.085 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-95% | 87% | 
| EDTA plasma(n=5) | 78-96% | 87% | 
| Heparin plasma(n=5) | 78-92% | 85% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 83-97% | 92-103% | 87-98% | 93-101% | 
| EDTA plasma(n=5) | 83-96% | 86-95% | 90-102% | 87-98% | 
| Heparin plasma(n=5) | 87-95% | 85-96% | 92-102% | 81-92% |