| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 0.182 | |
| 5.00 | 0.411 | |
| 2.50 | 0.715 | |
| 1.25 | 0.959 | |
| 0.63 | 1.253 | |
| 0.32 | 1.587 | |
| 0.16 | 1.761 | |
| 0.00 | 2.316 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 93-107% | 100% | 
| Heparin plasma(n=5) | 87-95% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-104% | 82-96% | 92-101% | 82-98% | 
| EDTA plasma(n=5) | 87-91% | 85-93% | 89-97% | 85-93% | 
| Heparin plasma(n=5) | 81-92% | 81-94% | 86-98% | 92-101% |