| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.093 | 2.007 |
| 100.00 | 1.553 | 1.467 |
| 50.00 | 1.112 | 1.026 |
| 25.00 | 0.847 | 0.761 |
| 12.50 | 0.526 | 0.440 |
| 6.25 | 0.369 | 0.283 |
| 3.13 | 0.191 | 0.105 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-116% | 103% |
| EDTA plasma(n=5) | 92-109% | 101% |
| Heparin plasma(n=5) | 83-100% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-118% | 84-117% | 85-94% | 81-99% |
| EDTA plasma(n=5) | 82-98% | 90-112% | 80-96% | 87-118% |
| Heparin plasma(n=5) | 81-100% | 83-98% | 87-115% | 87-119% |
