| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.051 | 1.966 |
| 5.00 | 1.606 | 1.521 |
| 2.50 | 1.211 | 1.126 |
| 1.25 | 0.809 | 0.724 |
| 0.63 | 0.496 | 0.411 |
| 0.31 | 0.318 | 0.233 |
| 0.16 | 0.215 | 0.130 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-119% | 104% |
| EDTA plasma(n=5) | 89-118% | 104% |
| Heparin plasma(n=5) | 86-110% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-115% | 91-102% | 81-100% | 83-93% |
| EDTA plasma(n=5) | 86-112% | 88-110% | 92-117% | 83-92% |
| Heparin plasma(n=5) | 92-115% | 82-111% | 94-113% | 94-120% |