| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.045 | 1.961 | 
| 5.00 | 1.593 | 1.509 | 
| 2.50 | 1.186 | 1.102 | 
| 1.25 | 0.828 | 0.744 | 
| 0.63 | 0.509 | 0.425 | 
| 0.32 | 0.325 | 0.241 | 
| 0.16 | 0.221 | 0.137 | 
| 0.00 | 0.084 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-95% | 88% | 
| EDTA plasma(n=5) | 78-92% | 85% | 
| Heparin plasma(n=5) | 88-106% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-102% | 96-104% | 87-98% | 95-107% | 
| EDTA plasma(n=5) | 87-97% | 86-92% | 89-103% | 89-98% | 
| Heparin plasma(n=5) | 79-85% | 87-96% | 82-95% | 89-103% |