| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.048 | 1.966 |
| 2000.00 | 1.602 | 1.520 |
| 1000.00 | 1.113 | 1.031 |
| 500.00 | 0.871 | 0.789 |
| 250.00 | 0.504 | 0.422 |
| 125.00 | 0.304 | 0.222 |
| 62.50 | 0.185 | 0.103 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-98% | 96% |
| EDTA plasma(n=5) | 91-124% | 108% |
| Heparin plasma(n=5) | 89-116% | 103% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-104% | 91-115% | 85-98% | 81-98% |
| EDTA plasma(n=5) | 87-120% | 86-107% | 84-95% | 91-106% |
| Heparin plasma(n=5) | 94-114% | 90-101% | 83-95% | 86-100% |