| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 2.154 | 2.061 | 
| 1000.00 | 1.642 | 1.549 | 
| 500.00 | 1.121 | 1.028 | 
| 250.00 | 0.875 | 0.782 | 
| 125.00 | 0.527 | 0.434 | 
| 62.50 | 0.326 | 0.233 | 
| 31.25 | 0.258 | 0.165 | 
| 0.00 | 0.093 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 92-105% | 98% | 
| EDTA plasma(n=5) | 83-95% | 89% | 
| Heparin plasma(n=5) | 79-93% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-102% | 82-94% | 85-96% | 78-90% | 
| EDTA plasma(n=5) | 93-102% | 80-93% | 91-100% | 88-95% | 
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |