| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.027 | 1.931 |
| 5.00 | 1.536 | 1.440 |
| 2.50 | 1.184 | 1.088 |
| 1.25 | 0.882 | 0.786 |
| 0.63 | 0.576 | 0.480 |
| 0.32 | 0.381 | 0.285 |
| 0.16 | 0.256 | 0.160 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-106% | 100% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 87-93% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-102% | 87-103% | 85-101% | 81-99% |
| EDTA plasma(n=5) | 95-104% | 93-105% | 82-101% | 91-103% |
| Heparin plasma(n=5) | 83-95% | 88-102% | 85-97% | 80-92% |