| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.916 | 1.834 |
| 500.00 | 1.545 | 1.463 |
| 250.00 | 1.223 | 1.141 |
| 125.00 | 0.847 | 0.765 |
| 62.50 | 0.507 | 0.425 |
| 31.25 | 0.357 | 0.275 |
| 15.63 | 0.257 | 0.175 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 82-97% | 89% |
| Heparin plasma(n=5) | 87-95% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-97% | 93-106% | 83-95% | 88-103% |
| EDTA plasma(n=5) | 95-103% | 88-96% | 81-95% | 91-102% |
| Heparin plasma(n=5) | 89-97% | 86-102% | 94-106% | 85-97% |