| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.054 | 1.961 |
| 2000.00 | 1.638 | 1.545 |
| 1000.00 | 1.143 | 1.050 |
| 500.00 | 0.878 | 0.785 |
| 250.00 | 0.509 | 0.416 |
| 125.00 | 0.357 | 0.264 |
| 62.50 | 0.186 | 0.093 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-91% | 84% |
| EDTA plasma(n=5) | 88-101% | 94% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-104% | 82-96% | 88-101% | 90-99% |
| EDTA plasma(n=5) | 88-97% | 95-106% | 83-98% | 93-105% |
| Heparin plasma(n=5) | 88-101% | 90-98% | 92-106% | 85-97% |