| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.909 | 1.827 |
| 5.00 | 1.551 | 1.469 |
| 2.50 | 1.210 | 1.128 |
| 1.25 | 0.869 | 0.787 |
| 0.63 | 0.486 | 0.404 |
| 0.31 | 0.382 | 0.300 |
| 0.16 | 0.253 | 0.171 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-109% | 98% |
| EDTA plasma(n=5) | 88-99% | 94% |
| Heparin plasma(n=5) | 84-113% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-119% | 87-101% | 91-112% | 87-117% |
| EDTA plasma(n=5) | 82-98% | 95-108% | 94-101% | 85-119% |
| Heparin plasma(n=5) | 90-106% | 88-100% | 94-107% | 92-115% |