| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.054 | 1.971 |
| 1000.00 | 1.528 | 1.445 |
| 500.00 | 1.282 | 1.199 |
| 250.00 | 0.928 | 0.845 |
| 125.00 | 0.563 | 0.480 |
| 62.50 | 0.389 | 0.306 |
| 31.25 | 0.216 | 0.133 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-95% | 88% |
| EDTA plasma(n=5) | 82-97% | 92% |
| Heparin plasma(n=5) | 87-99% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-97% | 87-96% | 85-96% | 85-95% |
| EDTA plasma(n=5) | 86-95% | 93-103% | 87-98% | 85-96% |
| Heparin plasma(n=5) | 92-102% | 81-93% | 86-98% | 91-103% |