| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.059 | 1.977 |
| 25.00 | 1.660 | 1.578 |
| 12.50 | 1.235 | 1.153 |
| 6.25 | 0.881 | 0.799 |
| 3.13 | 0.540 | 0.458 |
| 1.56 | 0.254 | 0.172 |
| 0.78 | 0.208 | 0.126 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 81-98% | 90% |
| Heparin plasma(n=5) | 84-90% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-105% | 94-107% | 91-102% | 92-110% |
| EDTA plasma(n=5) | 93-112% | 92-103% | 82-105% | 88-118% |
| Heparin plasma(n=5) | 90-99% | 81-112% | 85-109% | 93-107% |