| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.955 | 1.870 |
| 10.00 | 1.641 | 1.556 |
| 5.00 | 1.114 | 1.029 |
| 2.50 | 0.909 | 0.824 |
| 1.25 | 0.619 | 0.534 |
| 0.63 | 0.390 | 0.305 |
| 0.31 | 0.210 | 0.125 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-102% | 93% |
| EDTA plasma(n=5) | 82-88% | 85% |
| Heparin plasma(n=5) | 90-100% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 94-105% | 84-91% | 82-109% | 91-120% |
| EDTA plasma(n=5) | 87-105% | 90-117% | 85-109% | 82-110% |
| Heparin plasma(n=5) | 91-115% | 86-95% | 87-95% | 86-107% |