| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.925 | 1.842 |
| 5.00 | 1.538 | 1.455 |
| 2.50 | 1.181 | 1.098 |
| 1.25 | 0.867 | 0.784 |
| 0.63 | 0.541 | 0.458 |
| 0.31 | 0.241 | 0.158 |
| 0.16 | 0.185 | 0.102 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-105% | 94% |
| EDTA plasma(n=5) | 84-108% | 96% |
| Heparin plasma(n=5) | 93-119% | 106% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-100% | 94-111% | 93-118% | 92-115% |
| EDTA plasma(n=5) | 80-87% | 82-97% | 81-92% | 94-116% |
| Heparin plasma(n=5) | 93-116% | 87-117% | 95-103% | 92-113% |