| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 6000.00 | 2.095 | 1.992 |
| 3000.00 | 1.582 | 1.479 |
| 1500.00 | 1.151 | 1.048 |
| 750.00 | 0.905 | 0.802 |
| 375.00 | 0.538 | 0.435 |
| 187.50 | 0.317 | 0.214 |
| 93.75 | 0.219 | 0.116 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 84-96% | 90% |
| Heparin plasma(n=5) | 87-101% | 94% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-102% | 87-96% | 82-90% | 92-101% |
| EDTA plasma(n=5) | 79-92% | 85-96% | 80-98% | 95-102% |
| Heparin plasma(n=5) | 88-97% | 86-102% | 91-103% | 89-97% |