| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.270 | 2.185 |
| 5.00 | 1.559 | 1.474 |
| 2.50 | 1.173 | 1.088 |
| 1.25 | 0.765 | 0.680 |
| 0.63 | 0.611 | 0.526 |
| 0.32 | 0.329 | 0.244 |
| 0.16 | 0.216 | 0.131 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-97% | 90% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 77-95% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-104% | 79-92% | 82-95% | 86-97% |
| EDTA plasma(n=5) | 85-92% | 89-103% | 97-105% | 91-99% |
| Heparin plasma(n=5) | 88-102% | 86-92% | 81-93% | 89-104% |