| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 6000.00 | 0.227 | |
| 3000.00 | 0.397 | |
| 1500.00 | 0.611 | |
| 750.00 | 0.833 | |
| 375.00 | 1.157 | |
| 187.50 | 1.592 | |
| 93.75 | 1.715 | |
| 0.00 | 2.226 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 80-97% | 88% |
| Heparin plasma(n=5) | 88-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-101% | 87-98% | 85-94% | 87-101% |
| EDTA plasma(n=5) | 83-98% | 88-97% | 96-105% | 86-92% |
| Heparin plasma(n=5) | 79-92% | 85-96% | 87-102% | 91-99% |