| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.179 | 2.083 |
| 5.00 | 1.527 | 1.431 |
| 2.50 | 1.128 | 1.032 |
| 1.25 | 0.775 | 0.679 |
| 0.63 | 0.523 | 0.427 |
| 0.32 | 0.347 | 0.251 |
| 0.16 | 0.226 | 0.130 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-102% | 95% |
| EDTA plasma(n=5) | 79-97% | 88% |
| Heparin plasma(n=5) | 83-95% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-101% | 85-96% | 95-104% | 82-90% |
| EDTA plasma(n=5) | 95-104% | 97-106% | 85-98% | 81-92% |
| Heparin plasma(n=5) | 83-92% | 96-102% | 81-93% | 79-91% |