| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.054 | 1.958 |
| 2000.00 | 1.328 | 1.232 |
| 1000.00 | 1.106 | 1.010 |
| 500.00 | 0.836 | 0.740 |
| 250.00 | 0.794 | 0.698 |
| 125.00 | 0.342 | 0.246 |
| 62.50 | 0.157 | 0.061 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-93% | 86% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 85-95% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-99% | 82-96% | 92-101% | 82-90% |
| EDTA plasma(n=5) | 79-92% | 80-93% | 81-98% | 89-102% |
| Heparin plasma(n=5) | 89-97% | 86-95% | 81-93% | 91-103% |