| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.031 | 1.932 |
| 5.00 | 1.582 | 1.483 |
| 2.50 | 1.162 | 1.063 |
| 1.25 | 0.892 | 0.793 |
| 0.63 | 0.517 | 0.418 |
| 0.31 | 0.311 | 0.212 |
| 0.16 | 0.221 | 0.122 |
| 0.00 | 0.099 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-102% | 96% |
| EDTA plasma(n=5) | 82-124% | 103% |
| Heparin plasma(n=5) | 85-113% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-98% | 92-104% | 81-109% | 81-113% |
| EDTA plasma(n=5) | 81-88% | 85-99% | 94-117% | 87-98% |
| Heparin plasma(n=5) | 83-102% | 91-105% | 87-97% | 86-99% |