| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.172 | 2.085 |
| 1000.00 | 1.507 | 1.420 |
| 500.00 | 1.142 | 1.055 |
| 250.00 | 0.782 | 0.695 |
| 125.00 | 0.535 | 0.448 |
| 62.50 | 0.342 | 0.255 |
| 31.25 | 0.187 | 0.100 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 83-99% | 95% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-93% | 78-91% | 87-98% | 95-107% |
| EDTA plasma(n=5) | 78-92% | 79-93% | 94-106% | 83-97% |
| Heparin plasma(n=5) | 83-95% | 88-102% | 85-97% | 80-92% |