| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.943 | 1.846 |
| 2500.00 | 1.630 | 1.533 |
| 1250.00 | 1.185 | 1.088 |
| 625.00 | 0.808 | 0.711 |
| 312.50 | 0.508 | 0.411 |
| 156.25 | 0.406 | 0.309 |
| 78.13 | 0.211 | 0.114 |
| 0.00 | 0.097 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-110% | 100% |
| EDTA plasma(n=5) | 87-97% | 92% |
| Heparin plasma(n=5) | 85-109% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-118% | 89-100% | 84-114% | 83-96% |
| EDTA plasma(n=5) | 94-99% | 88-99% | 85-97% | 87-98% |
| Heparin plasma(n=5) | 83-109% | 87-108% | 93-98% | 90-115% |