| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.136 | 2.031 |
| 5.00 | 1.502 | 1.397 |
| 2.50 | 1.136 | 1.031 |
| 1.25 | 0.826 | 0.721 |
| 0.63 | 0.583 | 0.478 |
| 0.32 | 0.426 | 0.321 |
| 0.16 | 0.197 | 0.092 |
| 0.00 | 0.105 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-101% | 95% |
| EDTA plasma(n=5) | 82-95% | 88% |
| Heparin plasma(n=5) | 80-94% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-100% | 84-99% | 85-97% | 88-102% |
| EDTA plasma(n=5) | 86-102% | 87-96% | 89-98% | 83-103% |
| Heparin plasma(n=5) | 87-98% | 92-104% | 82-96% | 95-102% |