| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.265 | 2.169 |
| 5.00 | 1.690 | 1.594 |
| 2.50 | 1.162 | 1.066 |
| 1.25 | 0.754 | 0.658 |
| 0.63 | 0.582 | 0.486 |
| 0.31 | 0.335 | 0.239 |
| 0.16 | 0.225 | 0.129 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-113% | 99% |
| EDTA plasma(n=5) | 87-102% | 95% |
| Heparin plasma(n=5) | 84-95% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-101% | 83-114% | 88-95% | 88-111% |
| EDTA plasma(n=5) | 84-114% | 86-94% | 81-92% | 92-104% |
| Heparin plasma(n=5) | 89-96% | 92-101% | 91-117% | 84-98% |