| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.379 | 2.280 |
| 5.00 | 1.771 | 1.672 |
| 2.50 | 1.147 | 1.048 |
| 1.25 | 0.841 | 0.742 |
| 0.63 | 0.521 | 0.422 |
| 0.32 | 0.361 | 0.262 |
| 0.16 | 0.213 | 0.114 |
| 0.00 | 0.099 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-108% | 101% |
| EDTA plasma(n=5) | 83-97% | 90% |
| Heparin plasma(n=5) | 80-97% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 87-103% | 90-98% | 95-102% |
| EDTA plasma(n=5) | 79-93% | 86-92% | 84-98% | 89-102% |
| Heparin plasma(n=5) | 85-97% | 93-101% | 97-103% | 91-99% |