| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 1.958 | 1.857 |
| 2000.00 | 1.668 | 1.567 |
| 1000.00 | 1.190 | 1.089 |
| 500.00 | 0.919 | 0.818 |
| 250.00 | 0.543 | 0.442 |
| 125.00 | 0.326 | 0.225 |
| 62.50 | 0.269 | 0.168 |
| 0.00 | 0.101 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-116% | 100% |
| EDTA plasma(n=5) | 87-92% | 90% |
| Heparin plasma(n=5) | 88-108% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-109% | 87-107% | 95-114% | 91-108% |
| EDTA plasma(n=5) | 83-100% | 93-115% | 83-90% | 85-90% |
| Heparin plasma(n=5) | 93-117% | 94-111% | 94-109% | 81-99% |