| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.911 | 1.818 |
| 5.00 | 1.560 | 1.467 |
| 2.50 | 1.121 | 1.028 |
| 1.25 | 0.871 | 0.778 |
| 0.63 | 0.524 | 0.431 |
| 0.31 | 0.357 | 0.264 |
| 0.16 | 0.183 | 0.090 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-106% | 97% |
| EDTA plasma(n=5) | 92-107% | 100% |
| Heparin plasma(n=5) | 84-106% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-118% | 88-109% | 87-107% | 84-91% |
| EDTA plasma(n=5) | 92-107% | 84-111% | 89-104% | 82-92% |
| Heparin plasma(n=5) | 82-107% | 94-115% | 94-106% | 83-92% |