| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.219 | 2.116 | 
| 5.00 | 1.700 | 1.597 | 
| 2.50 | 1.168 | 1.065 | 
| 1.25 | 0.845 | 0.742 | 
| 0.63 | 0.508 | 0.405 | 
| 0.32 | 0.318 | 0.215 | 
| 0.16 | 0.174 | 0.071 | 
| 0.00 | 0.103 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 78-90% | 84% | 
| Heparin plasma(n=5) | 89-107% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 82-94% | 85-96% | 78-90% | 91-99% | 
| EDTA plasma(n=5) | 79-95% | 96-105% | 85-96% | 90-99% | 
| Heparin plasma(n=5) | 78-92% | 83-96% | 95-107% | 87-102% |