Products

Human FLNb(Filamin B Beta) ELISA Kit
One-click to copy product information
ELK3056
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Human FLNb(Filamin B Beta) ELISA Kit
Reactivity: Human
Alternative Names: ABP-278; AOI; FH1; FLN1L; LRS1; SCT; TABP; TAP; Actin Binding Protein 278; Actin-binding-like protein; Thyroid autoantigen; Filamin-3; Truncated actin-binding protein
Assay Type: Sandwich
Sensitivity: 5.2 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample type: Tissue homogenates, cell lysates and other biological fluids.
Assay length: 3.5h
Research Area: Developmental science;
Uniprot ID: O75369
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human FLNb. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human FLNb. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human FLNb, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human FLNb in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
1000.00 2.252 2.181
500.00 1.699 1.628
250.00 1.135 1.064
125.00 0.864 0.793
62.50 0.496 0.425
31.25 0.388 0.317
15.63 0.235 0.164
0.00 0.071 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant FLNb and the recovery rates were calculated by comparing the measured value to the expected amount of FLNb in samples.
Matrix Recovery range Average
serum(n=5) 80-95% 87%
EDTA plasma(n=5) 80-95% 87%
Heparin plasma(n=5) 86-99% 92%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of FLNb and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 81-94% 96-105% 92-103% 82-96%
EDTA plasma(n=5) 82-94% 85-96% 78-90% 91-99%
Heparin plasma(n=5) 86-93% 91-105% 87-98% 99-102%