| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 2.185 | 2.099 | 
| 1000.00 | 1.652 | 1.566 | 
| 500.00 | 1.223 | 1.137 | 
| 250.00 | 0.854 | 0.768 | 
| 125.00 | 0.555 | 0.469 | 
| 62.50 | 0.303 | 0.217 | 
| 31.25 | 0.175 | 0.089 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-95% | 91% | 
| EDTA plasma(n=5) | 78-90% | 84% | 
| Heparin plasma(n=5) | 84-97% | 90% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-93% | 93-101% | 86-97% | 90-101% | 
| EDTA plasma(n=5) | 86-96% | 91-107% | 87-101% | 95-102% | 
| Heparin plasma(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |