| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.138 | 2.051 | 
| 5.00 | 1.588 | 1.501 | 
| 2.50 | 1.106 | 1.019 | 
| 1.25 | 0.958 | 0.871 | 
| 0.63 | 0.623 | 0.536 | 
| 0.32 | 0.384 | 0.297 | 
| 0.16 | 0.256 | 0.169 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 91-105% | 98% | 
| EDTA plasma(n=5) | 96-104% | 100% | 
| Heparin plasma(n=5) | 82-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 93-105% | 82-101% | 91-103% | 95-102% | 
| EDTA plasma(n=5) | 88-104% | 84-98% | 95-105% | 88-102% | 
| Heparin plasma(n=5) | 92-107% | 83-99% | 80-96% | 91-101% |