| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.064 | 1.957 |
| 250.00 | 1.538 | 1.431 |
| 125.00 | 1.119 | 1.012 |
| 62.50 | 0.933 | 0.826 |
| 31.25 | 0.602 | 0.495 |
| 15.63 | 0.317 | 0.210 |
| 7.81 | 0.181 | 0.074 |
| 0.00 | 0.107 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 91-122% | 107% |
| Heparin plasma(n=5) | 94-116% | 105% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-115% | 84-119% | 87-98% | 89-113% |
| EDTA plasma(n=5) | 88-99% | 88-118% | 92-107% | 80-118% |
| Heparin plasma(n=5) | 87-114% | 92-119% | 90-118% | 88-95% |