| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 2.266 | 2.172 | 
| 250.00 | 1.613 | 1.519 | 
| 125.00 | 1.115 | 1.021 | 
| 62.50 | 0.705 | 0.611 | 
| 31.25 | 0.589 | 0.495 | 
| 15.63 | 0.297 | 0.203 | 
| 7.82 | 0.216 | 0.122 | 
| 0.00 | 0.094 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-97% | 88% | 
| EDTA plasma(n=5) | 78-92% | 85% | 
| Heparin plasma(n=5) | 90-105% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-96% | 91-107% | 87-101% | 95-102% | 
| EDTA plasma(n=5) | 89-102% | 92-101% | 88-102% | 87-102% | 
| Heparin plasma(n=5) | 85-97% | 84-100% | 95-103% | 88-104% |