| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 1.842 | 1.755 | 
| 2500.00 | 1.447 | 1.360 | 
| 1250.00 | 1.186 | 1.099 | 
| 625.00 | 0.883 | 0.796 | 
| 312.50 | 0.574 | 0.487 | 
| 156.25 | 0.388 | 0.301 | 
| 78.13 | 0.211 | 0.124 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-95% | 87% | 
| EDTA plasma(n=5) | 87-99% | 93% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-94% | 90-99% | 95-102% | 82-90% | 
| EDTA plasma(n=5) | 89-98% | 83-103% | 98-105% | 92-101% | 
| Heparin plasma(n=5) | 90-99% | 83-97% | 81-103% | 89-101% |