| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.075 | 2.010 | 
| 2500.00 | 1.502 | 1.437 | 
| 1250.00 | 0.939 | 0.874 | 
| 625.00 | 0.747 | 0.682 | 
| 312.50 | 0.546 | 0.481 | 
| 156.25 | 0.312 | 0.247 | 
| 78.13 | 0.217 | 0.152 | 
| 0.00 | 0.065 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-92% | 85% | 
| EDTA plasma(n=5) | 78-90% | 84% | 
| Heparin plasma(n=5) | 85-99% | 92% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 78-92% | 79-93% | 94-106% | 83-97% | 
| EDTA plasma(n=5) | 80-101% | 87-96% | 96-105% | 95-103% | 
| Heparin plasma(n=5) | 87-96% | 91-101% | 87-98% | 89-103% |