| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.245 | 2.164 | 
| 500.00 | 1.697 | 1.616 | 
| 250.00 | 1.151 | 1.070 | 
| 125.00 | 0.752 | 0.671 | 
| 62.50 | 0.542 | 0.461 | 
| 31.25 | 0.425 | 0.344 | 
| 15.63 | 0.152 | 0.071 | 
| 0.00 | 0.081 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 88-102% | 95% | 
| EDTA plasma(n=5) | 80-92% | 86% | 
| Heparin plasma(n=5) | 90-105% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-93% | 91-101% | 87-98% | 95-104% | 
| EDTA plasma(n=5) | 86-94% | 93-102% | 89-97% | 85-94% | 
| Heparin plasma(n=5) | 85-94% | 87-101% | 89-99% | 85-95% |