| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.096 | 2.011 |
| 250.00 | 1.446 | 1.361 |
| 125.00 | 0.835 | 0.750 |
| 62.50 | 0.702 | 0.617 |
| 31.25 | 0.481 | 0.396 |
| 15.63 | 0.330 | 0.245 |
| 7.81 | 0.189 | 0.104 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-93% | 89% |
| EDTA plasma(n=5) | 90-117% | 104% |
| Heparin plasma(n=5) | 88-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-119% | 86-104% | 92-111% | 86-110% |
| EDTA plasma(n=5) | 87-109% | 90-119% | 95-105% | 90-116% |
| Heparin plasma(n=5) | 81-112% | 83-115% | 93-115% | 90-97% |