| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.413 | 2.323 |
| 5.00 | 1.646 | 1.556 |
| 2.50 | 0.946 | 0.856 |
| 1.25 | 0.687 | 0.597 |
| 0.63 | 0.450 | 0.360 |
| 0.31 | 0.316 | 0.226 |
| 0.16 | 0.141 | 0.051 |
| 0.00 | 0.090 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-115% | 104% |
| EDTA plasma(n=5) | 94-102% | 98% |
| Heparin plasma(n=5) | 82-111% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-100% | 88-97% | 83-107% | 84-114% |
| EDTA plasma(n=5) | 93-103% | 91-100% | 91-98% | 89-110% |
| Heparin plasma(n=5) | 81-93% | 90-97% | 82-90% | 80-92% |