| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 15.00 | 2.054 | 1.956 | 
| 7.50 | 1.496 | 1.398 | 
| 3.75 | 1.142 | 1.044 | 
| 1.88 | 0.775 | 0.677 | 
| 0.94 | 0.594 | 0.496 | 
| 0.47 | 0.423 | 0.325 | 
| 0.24 | 0.214 | 0.116 | 
| 0.00 | 0.098 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-94% | 88% | 
| EDTA plasma(n=5) | 81-97% | 89% | 
| Heparin plasma(n=5) | 84-97% | 90% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 92-101% | 82-98% | 89-97% | 85-93% | 
| EDTA plasma(n=5) | 87-94% | 85-96% | 89-100% | 85-96% | 
| Heparin plasma(n=5) | 87-96% | 85-97% | 85-94% | 87-96% |