| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 1.952 | 1.883 | 
| 5.00 | 1.674 | 1.605 | 
| 2.50 | 1.168 | 1.099 | 
| 1.25 | 0.797 | 0.728 | 
| 0.63 | 0.532 | 0.463 | 
| 0.32 | 0.340 | 0.271 | 
| 0.16 | 0.196 | 0.127 | 
| 0.00 | 0.069 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 90-104% | 97% | 
| EDTA plasma(n=5) | 80-97% | 88% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-98% | 87-100% | 85-94% | 83-96% | 
| EDTA plasma(n=5) | 95-106% | 83-96% | 87-98% | 78-89% | 
| Heparin plasma(n=5) | 87-96% | 82-95% | 93-102% | 87-96% |